Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: In phantom feasibility study

G. Gambarini, M. Carrara, C. Tenconi, N. Mantaut, M. Borroni, D. Cutajar, M. Petasecca, I. Fuduli, M. Lerch, E. Pignoli, A. Rosenfeld

Research output: Contribution to journalArticlepeer-review

Abstract

MO. Skin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MO. Skin and dual-MO. Skin configurations, respectively. Results show that dual-MO. Skin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.

Original languageEnglish
Pages (from-to)222-226
Number of pages5
JournalApplied Radiation and Isotopes
Volume83
DOIs
Publication statusPublished - Jan 2014

Keywords

  • High dose rate
  • In vivo dosimetry
  • MOSFET
  • Prostate brachytherapy

ASJC Scopus subject areas

  • Radiation

Fingerprint Dive into the research topics of 'Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: In phantom feasibility study'. Together they form a unique fingerprint.