Optimised procedure for the calibration of the force platform location

M. Rabuffetti, M. Ferrarin, P. Mazzoleni, F. Benvenuti, A. Pedotti

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

An innovative optimised method, including an experiment and a mathematical model, for the calibration of the force platform location in the optoelectronic reference frame is proposed. The calibration experiment adopts a bearing-marker testing object contacting the platform and does not directly measure the platform location. The experiment is designed in order to avoid the main drawbacks possibly occurring in commonly adopted methods. The mathematical model of the experiment estimates the expected ground reaction. An optimisation algorithm identifies the optimal platform location as the one that best matches the measured outcome of the calibration experiment with the corresponding model estimate. The innovative calibration procedure has been assessed in terms of inter-tester reliability and compared with commonly used calibration procedures of platform location. These results evidenced how the introduction of such optimised procedure could improve the reliability of the calibrated platform location and, consequently, of the kinetic variables considered in posture and gait analysis.

Original languageEnglish
Pages (from-to)75-80
Number of pages6
JournalGait and Posture
Volume17
Issue number1
DOIs
Publication statusPublished - Feb 2003

Fingerprint

Calibration
Theoretical Models
Posture
Gait

Keywords

  • Calibration
  • Force platform
  • Laboratory management
  • Motion analysis
  • Optimisation
  • Rehabilitation outcome assessment

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Optimised procedure for the calibration of the force platform location. / Rabuffetti, M.; Ferrarin, M.; Mazzoleni, P.; Benvenuti, F.; Pedotti, A.

In: Gait and Posture, Vol. 17, No. 1, 02.2003, p. 75-80.

Research output: Contribution to journalArticle

Rabuffetti, M. ; Ferrarin, M. ; Mazzoleni, P. ; Benvenuti, F. ; Pedotti, A. / Optimised procedure for the calibration of the force platform location. In: Gait and Posture. 2003 ; Vol. 17, No. 1. pp. 75-80.
@article{19cba8583cbb49b9917b061c0e60ccbc,
title = "Optimised procedure for the calibration of the force platform location",
abstract = "An innovative optimised method, including an experiment and a mathematical model, for the calibration of the force platform location in the optoelectronic reference frame is proposed. The calibration experiment adopts a bearing-marker testing object contacting the platform and does not directly measure the platform location. The experiment is designed in order to avoid the main drawbacks possibly occurring in commonly adopted methods. The mathematical model of the experiment estimates the expected ground reaction. An optimisation algorithm identifies the optimal platform location as the one that best matches the measured outcome of the calibration experiment with the corresponding model estimate. The innovative calibration procedure has been assessed in terms of inter-tester reliability and compared with commonly used calibration procedures of platform location. These results evidenced how the introduction of such optimised procedure could improve the reliability of the calibrated platform location and, consequently, of the kinetic variables considered in posture and gait analysis.",
keywords = "Calibration, Force platform, Laboratory management, Motion analysis, Optimisation, Rehabilitation outcome assessment",
author = "M. Rabuffetti and M. Ferrarin and P. Mazzoleni and F. Benvenuti and A. Pedotti",
year = "2003",
month = "2",
doi = "10.1016/S0966-6362(02)00061-9",
language = "English",
volume = "17",
pages = "75--80",
journal = "Gait and Posture",
issn = "0966-6362",
publisher = "Elsevier B.V.",
number = "1",

}

TY - JOUR

T1 - Optimised procedure for the calibration of the force platform location

AU - Rabuffetti, M.

AU - Ferrarin, M.

AU - Mazzoleni, P.

AU - Benvenuti, F.

AU - Pedotti, A.

PY - 2003/2

Y1 - 2003/2

N2 - An innovative optimised method, including an experiment and a mathematical model, for the calibration of the force platform location in the optoelectronic reference frame is proposed. The calibration experiment adopts a bearing-marker testing object contacting the platform and does not directly measure the platform location. The experiment is designed in order to avoid the main drawbacks possibly occurring in commonly adopted methods. The mathematical model of the experiment estimates the expected ground reaction. An optimisation algorithm identifies the optimal platform location as the one that best matches the measured outcome of the calibration experiment with the corresponding model estimate. The innovative calibration procedure has been assessed in terms of inter-tester reliability and compared with commonly used calibration procedures of platform location. These results evidenced how the introduction of such optimised procedure could improve the reliability of the calibrated platform location and, consequently, of the kinetic variables considered in posture and gait analysis.

AB - An innovative optimised method, including an experiment and a mathematical model, for the calibration of the force platform location in the optoelectronic reference frame is proposed. The calibration experiment adopts a bearing-marker testing object contacting the platform and does not directly measure the platform location. The experiment is designed in order to avoid the main drawbacks possibly occurring in commonly adopted methods. The mathematical model of the experiment estimates the expected ground reaction. An optimisation algorithm identifies the optimal platform location as the one that best matches the measured outcome of the calibration experiment with the corresponding model estimate. The innovative calibration procedure has been assessed in terms of inter-tester reliability and compared with commonly used calibration procedures of platform location. These results evidenced how the introduction of such optimised procedure could improve the reliability of the calibrated platform location and, consequently, of the kinetic variables considered in posture and gait analysis.

KW - Calibration

KW - Force platform

KW - Laboratory management

KW - Motion analysis

KW - Optimisation

KW - Rehabilitation outcome assessment

UR - http://www.scopus.com/inward/record.url?scp=0037306981&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037306981&partnerID=8YFLogxK

U2 - 10.1016/S0966-6362(02)00061-9

DO - 10.1016/S0966-6362(02)00061-9

M3 - Article

C2 - 12535729

AN - SCOPUS:0037306981

VL - 17

SP - 75

EP - 80

JO - Gait and Posture

JF - Gait and Posture

SN - 0966-6362

IS - 1

ER -