Performance evaluation of reflectance meter for glucose determination by two different reagent strips

Donatella Spotti, Cristina Rocco, Orazio Carandente

Research output: Contribution to journalArticle

Abstract

Serum glucose concentrations of 112 blood samples determined by the GOD/POD/Trinder method were compared with values obtained on whole blood by means of the Glucometer reflectance meter and two different reagent strips, Dextrostix and an experimental strip (GX 947822), in order to establish over a wide range of glucose concentrations, the precision and reproducibility of reflectometric methods. The two methods examined showed an excellent correlation with the reference method, particularly if data were corrected for the individual hematocrit value, and both accuracy and precision were reasonably satisfactory.

Original languageEnglish
Pages (from-to)149-158
Number of pages10
JournalActa Diabetologica Latina
Volume22
Issue number2
DOIs
Publication statusPublished - Apr 1985

Fingerprint

Reagent Strips
Glucose
Hematocrit
Serum

Keywords

  • Diabetes
  • Glucose control
  • Reflectance meter
  • Self-monitoring

ASJC Scopus subject areas

  • Internal Medicine
  • Endocrinology, Diabetes and Metabolism
  • Endocrinology

Cite this

Performance evaluation of reflectance meter for glucose determination by two different reagent strips. / Spotti, Donatella; Rocco, Cristina; Carandente, Orazio.

In: Acta Diabetologica Latina, Vol. 22, No. 2, 04.1985, p. 149-158.

Research output: Contribution to journalArticle

Spotti, Donatella ; Rocco, Cristina ; Carandente, Orazio. / Performance evaluation of reflectance meter for glucose determination by two different reagent strips. In: Acta Diabetologica Latina. 1985 ; Vol. 22, No. 2. pp. 149-158.
@article{06b4cc7c26ba4dce86b691391e50783a,
title = "Performance evaluation of reflectance meter for glucose determination by two different reagent strips",
abstract = "Serum glucose concentrations of 112 blood samples determined by the GOD/POD/Trinder method were compared with values obtained on whole blood by means of the Glucometer reflectance meter and two different reagent strips, Dextrostix and an experimental strip (GX 947822), in order to establish over a wide range of glucose concentrations, the precision and reproducibility of reflectometric methods. The two methods examined showed an excellent correlation with the reference method, particularly if data were corrected for the individual hematocrit value, and both accuracy and precision were reasonably satisfactory.",
keywords = "Diabetes, Glucose control, Reflectance meter, Self-monitoring",
author = "Donatella Spotti and Cristina Rocco and Orazio Carandente",
year = "1985",
month = "4",
doi = "10.1007/BF02590789",
language = "English",
volume = "22",
pages = "149--158",
journal = "Acta Diabetologica",
issn = "0940-5429",
publisher = "Springer-Verlag Italia s.r.l.",
number = "2",

}

TY - JOUR

T1 - Performance evaluation of reflectance meter for glucose determination by two different reagent strips

AU - Spotti, Donatella

AU - Rocco, Cristina

AU - Carandente, Orazio

PY - 1985/4

Y1 - 1985/4

N2 - Serum glucose concentrations of 112 blood samples determined by the GOD/POD/Trinder method were compared with values obtained on whole blood by means of the Glucometer reflectance meter and two different reagent strips, Dextrostix and an experimental strip (GX 947822), in order to establish over a wide range of glucose concentrations, the precision and reproducibility of reflectometric methods. The two methods examined showed an excellent correlation with the reference method, particularly if data were corrected for the individual hematocrit value, and both accuracy and precision were reasonably satisfactory.

AB - Serum glucose concentrations of 112 blood samples determined by the GOD/POD/Trinder method were compared with values obtained on whole blood by means of the Glucometer reflectance meter and two different reagent strips, Dextrostix and an experimental strip (GX 947822), in order to establish over a wide range of glucose concentrations, the precision and reproducibility of reflectometric methods. The two methods examined showed an excellent correlation with the reference method, particularly if data were corrected for the individual hematocrit value, and both accuracy and precision were reasonably satisfactory.

KW - Diabetes

KW - Glucose control

KW - Reflectance meter

KW - Self-monitoring

UR - http://www.scopus.com/inward/record.url?scp=0021994832&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021994832&partnerID=8YFLogxK

U2 - 10.1007/BF02590789

DO - 10.1007/BF02590789

M3 - Article

C2 - 4072566

AN - SCOPUS:0021994832

VL - 22

SP - 149

EP - 158

JO - Acta Diabetologica

JF - Acta Diabetologica

SN - 0940-5429

IS - 2

ER -