T1 - Quantification of soft tissue artefact in motion analysis by combining 3D fluoroscopy and stereophotogrammetry
T2 - A study on two subjects
AU - Stagni, Rita
AU - Fantozzi, Silvia
AU - Cappello, Angelo
AU - Leardini, Alberto
PY - 2005/3
Y1 - 2005/3
N2 - Background. Soft tissue artefact is the most invalidating source of error in human motion analysis using optoelectronic stereophotogrammetry. It is caused by the erroneous assumption that markers attached to the skin surface are rigidly connected to the underlying bones. The quantification of this artefact in three dimensions and the knowledge of how it propagates to relevant joint angles is necessary for the interpretation of gait analysis data. Methods. Two subjects, treated by total knee replacement, underwent data acquisition simultaneously with fluoroscopy and stereophotogrammetry during stair climbing, step up/down, sit-to-stand/stand-to-sit, and extension against gravity. The reference 3D kinematics of the femur and tibia was reconstructed from fluoroscopy-based tracking of the relevant prosthesis components. Soft tissue artefact was quantified as the motion of a grid of retro-reflecting makers attached to the thigh and shank with respect to the underlying bones, tracked by optoelectronic stereophotogrammetry. The propagation of soft tissue artefact to knee rotations was also calculated. Findings. The standard deviation of skin marker trajectory in the corresponding prosthesis-embedded anatomical frame was found up to 31 mm for the thigh and up to 21 mm for the shank. The ab/adduction and internal/external rotation angles were the most affected by soft tissue artefact propagation, with root mean square errors up to 192% and 117% of the corresponding range, respectively. Interpretations. In both the analysed subjects the proximal thigh showed the largest soft tissue artefact. This is subject- and task-specific. However, larger artefact does not necessarily produce larger propagated error on knee rotations. Propagated errors were extremely critical on ab/adduction and internal/external rotation. These large errors can nullify the usefulness of these variables in the clinical interpretation of gait analysis.
AB - Background. Soft tissue artefact is the most invalidating source of error in human motion analysis using optoelectronic stereophotogrammetry. It is caused by the erroneous assumption that markers attached to the skin surface are rigidly connected to the underlying bones. The quantification of this artefact in three dimensions and the knowledge of how it propagates to relevant joint angles is necessary for the interpretation of gait analysis data. Methods. Two subjects, treated by total knee replacement, underwent data acquisition simultaneously with fluoroscopy and stereophotogrammetry during stair climbing, step up/down, sit-to-stand/stand-to-sit, and extension against gravity. The reference 3D kinematics of the femur and tibia was reconstructed from fluoroscopy-based tracking of the relevant prosthesis components. Soft tissue artefact was quantified as the motion of a grid of retro-reflecting makers attached to the thigh and shank with respect to the underlying bones, tracked by optoelectronic stereophotogrammetry. The propagation of soft tissue artefact to knee rotations was also calculated. Findings. The standard deviation of skin marker trajectory in the corresponding prosthesis-embedded anatomical frame was found up to 31 mm for the thigh and up to 21 mm for the shank. The ab/adduction and internal/external rotation angles were the most affected by soft tissue artefact propagation, with root mean square errors up to 192% and 117% of the corresponding range, respectively. Interpretations. In both the analysed subjects the proximal thigh showed the largest soft tissue artefact. This is subject- and task-specific. However, larger artefact does not necessarily produce larger propagated error on knee rotations. Propagated errors were extremely critical on ab/adduction and internal/external rotation. These large errors can nullify the usefulness of these variables in the clinical interpretation of gait analysis.
KW - Fluoroscopy
KW - Human motion analysis
KW - Knee kinematics
KW - Soft tissue artefact
KW - Stereophotogrammetry
UR - http://www.scopus.com/inward/record.url?scp=13844256386&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=13844256386&partnerID=8YFLogxK
U2 - 10.1016/j.clinbiomech.2004.11.012
DO - 10.1016/j.clinbiomech.2004.11.012
M3 - Article
C2 - 15698706
AN - SCOPUS:13844256386
VL - 20
SP - 320
EP - 329
JO - Clinical Biomechanics
JF - Clinical Biomechanics
SN - 0268-0033
IS - 3
ER -