Raman characterisation of conventional and cross-linked polyethylene in acetabular cups run on a hip joint simulator

P. Taddei, M. Di Foggia, S. Affatato

Research output: Contribution to journalArticle

17 Citations (Scopus)

Abstract

Five sets of differently sterilised conventional ultra-high molecular weight polyethylene (UHMWPE) and cross-linked polyethylene (XLPE) acetabular cups were run for 5 million cycles on a hip joint simulator in order to evaluate their wear behaviour in relation to material properties (PE grade, conventional or cross-linked) and sterilisation method (ethylene oxide (EtO) treatment or γ-irradiation). Gravimetric measurements revealed that conventional UHMWPE wore significantly more than XLPE. The differences in wear behaviour could be partly related to the orthorhombic contents obtained by Raman spectroscopy in the unworn areas of the cups: XLPE cups showed a significantly higher crystallinity degree than the UHMWPE specimens. Raman analysis showed that wear testing did not significantly modify the orthorhombic content of any of the tested acetabular cups. However, the set of cups that showed the highest weight loss, i.e. γ-sterilised PE GUR1020, appeared the most homogeneously polished upon wear testing; from a molecular point of view, only this set of cups showed a significant increase of the I1130/I1060 intensity ratio, suggesting the occurrence of chain orientation. On the other hand, XLPE cups, despite the lowest weight loss undergone, showed a decrease in the amorphous content upon wear testing as well as a limited orthorhombic → monoclinic transformation, which did not appear detrimental.

Original languageEnglish
Pages (from-to)1344-1352
Number of pages9
JournalJournal of Raman Spectroscopy
Volume42
Issue number6
DOIs
Publication statusPublished - Jun 2011

Fingerprint

Polyethylene
Polyethylenes
Simulators
Wear of materials
Ultrahigh molecular weight polyethylenes
Testing
Ethylene Oxide
Raman spectroscopy
Materials properties
Ethylene
Irradiation
Oxides
ultra-high molecular weight polyethylene

Keywords

  • conventional ultra-high molecular weight polyethylene
  • cross-linked polyethylene
  • hip joint simulator
  • micro-Raman spectroscopy
  • sterilisation

ASJC Scopus subject areas

  • Spectroscopy
  • Materials Science(all)

Cite this

Raman characterisation of conventional and cross-linked polyethylene in acetabular cups run on a hip joint simulator. / Taddei, P.; Di Foggia, M.; Affatato, S.

In: Journal of Raman Spectroscopy, Vol. 42, No. 6, 06.2011, p. 1344-1352.

Research output: Contribution to journalArticle

@article{08979de504334e30885e4f3e0eccf7c9,
title = "Raman characterisation of conventional and cross-linked polyethylene in acetabular cups run on a hip joint simulator",
abstract = "Five sets of differently sterilised conventional ultra-high molecular weight polyethylene (UHMWPE) and cross-linked polyethylene (XLPE) acetabular cups were run for 5 million cycles on a hip joint simulator in order to evaluate their wear behaviour in relation to material properties (PE grade, conventional or cross-linked) and sterilisation method (ethylene oxide (EtO) treatment or γ-irradiation). Gravimetric measurements revealed that conventional UHMWPE wore significantly more than XLPE. The differences in wear behaviour could be partly related to the orthorhombic contents obtained by Raman spectroscopy in the unworn areas of the cups: XLPE cups showed a significantly higher crystallinity degree than the UHMWPE specimens. Raman analysis showed that wear testing did not significantly modify the orthorhombic content of any of the tested acetabular cups. However, the set of cups that showed the highest weight loss, i.e. γ-sterilised PE GUR1020, appeared the most homogeneously polished upon wear testing; from a molecular point of view, only this set of cups showed a significant increase of the I1130/I1060 intensity ratio, suggesting the occurrence of chain orientation. On the other hand, XLPE cups, despite the lowest weight loss undergone, showed a decrease in the amorphous content upon wear testing as well as a limited orthorhombic → monoclinic transformation, which did not appear detrimental.",
keywords = "conventional ultra-high molecular weight polyethylene, cross-linked polyethylene, hip joint simulator, micro-Raman spectroscopy, sterilisation",
author = "P. Taddei and {Di Foggia}, M. and S. Affatato",
year = "2011",
month = "6",
doi = "10.1002/jrs.2867",
language = "English",
volume = "42",
pages = "1344--1352",
journal = "Journal of Raman Spectroscopy",
issn = "0377-0486",
publisher = "John Wiley and Sons Ltd",
number = "6",

}

TY - JOUR

T1 - Raman characterisation of conventional and cross-linked polyethylene in acetabular cups run on a hip joint simulator

AU - Taddei, P.

AU - Di Foggia, M.

AU - Affatato, S.

PY - 2011/6

Y1 - 2011/6

N2 - Five sets of differently sterilised conventional ultra-high molecular weight polyethylene (UHMWPE) and cross-linked polyethylene (XLPE) acetabular cups were run for 5 million cycles on a hip joint simulator in order to evaluate their wear behaviour in relation to material properties (PE grade, conventional or cross-linked) and sterilisation method (ethylene oxide (EtO) treatment or γ-irradiation). Gravimetric measurements revealed that conventional UHMWPE wore significantly more than XLPE. The differences in wear behaviour could be partly related to the orthorhombic contents obtained by Raman spectroscopy in the unworn areas of the cups: XLPE cups showed a significantly higher crystallinity degree than the UHMWPE specimens. Raman analysis showed that wear testing did not significantly modify the orthorhombic content of any of the tested acetabular cups. However, the set of cups that showed the highest weight loss, i.e. γ-sterilised PE GUR1020, appeared the most homogeneously polished upon wear testing; from a molecular point of view, only this set of cups showed a significant increase of the I1130/I1060 intensity ratio, suggesting the occurrence of chain orientation. On the other hand, XLPE cups, despite the lowest weight loss undergone, showed a decrease in the amorphous content upon wear testing as well as a limited orthorhombic → monoclinic transformation, which did not appear detrimental.

AB - Five sets of differently sterilised conventional ultra-high molecular weight polyethylene (UHMWPE) and cross-linked polyethylene (XLPE) acetabular cups were run for 5 million cycles on a hip joint simulator in order to evaluate their wear behaviour in relation to material properties (PE grade, conventional or cross-linked) and sterilisation method (ethylene oxide (EtO) treatment or γ-irradiation). Gravimetric measurements revealed that conventional UHMWPE wore significantly more than XLPE. The differences in wear behaviour could be partly related to the orthorhombic contents obtained by Raman spectroscopy in the unworn areas of the cups: XLPE cups showed a significantly higher crystallinity degree than the UHMWPE specimens. Raman analysis showed that wear testing did not significantly modify the orthorhombic content of any of the tested acetabular cups. However, the set of cups that showed the highest weight loss, i.e. γ-sterilised PE GUR1020, appeared the most homogeneously polished upon wear testing; from a molecular point of view, only this set of cups showed a significant increase of the I1130/I1060 intensity ratio, suggesting the occurrence of chain orientation. On the other hand, XLPE cups, despite the lowest weight loss undergone, showed a decrease in the amorphous content upon wear testing as well as a limited orthorhombic → monoclinic transformation, which did not appear detrimental.

KW - conventional ultra-high molecular weight polyethylene

KW - cross-linked polyethylene

KW - hip joint simulator

KW - micro-Raman spectroscopy

KW - sterilisation

UR - http://www.scopus.com/inward/record.url?scp=79959425842&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79959425842&partnerID=8YFLogxK

U2 - 10.1002/jrs.2867

DO - 10.1002/jrs.2867

M3 - Article

VL - 42

SP - 1344

EP - 1352

JO - Journal of Raman Spectroscopy

JF - Journal of Raman Spectroscopy

SN - 0377-0486

IS - 6

ER -