Recent advances and future perspectives of gamma imagers for scintimammography

R. Pani, R. Pellegrini, M. N. Cinti, P. Bennati, M. Betti, V. Casali, O. Schillaci, M. Mattioli, V. Orsolini Cencelli, F. Navarria, D. Bollini, G. Moschini, F. Garibaldi, F. Cusanno, G. Iurlaro, L. Montani, R. Scafè, F. de Notaristefani

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

The very low sensitivity of Scintimammography for tumors under 1 cm in diameter, with current nuclear medicine cameras in use, is the major limitation in recommending this test modality for screening purposes. Recently latest generation cameras with superior imaging performances have allowed to foresee a very promising future for scintimammography. Recent technological advances obtained from our research group on the new Lanthanum scintillation crystals are now demonstrating how continuous crystals coupled to Hamamatsu Flat panel tube can overcome a number of limitations in spatial and energy resolution of scintillation arrays, carrying out 6% energy resolution, and submillimeter spatial resolution values at 140 keV. In this paper three basic detection systems have been analysed: a LumaGEM 3200S gamma camera (Gamma Medica, InC., Northridge, USA), based on a cadmium-zinc-telluride (CZT) semiconductor detector with a FoV of 12.5×12.5 cm2 and with detector elements of 1.6×1.6 mm2 in size and a detector prototype consisting of a Flat panel PMT coupled to planar LaBr3:Ce scintillator. The third one is a first generation large FoV gamma camera chosen for comparison, based on PSPMTs photodetector array coupled to a NaI(Tl) scintillation crystal matrix. LaBr3:Ce and CZT cameras showed superior spatial and energy resolution than previous generation one based on NaI(Tl) scintillation array. Both CZT and LaBr gamma cameras showed similar efficiencies for the same energy window though LaBr3:Ce continuous crystal showed better imaging performance than pixellated detectors. However large area LaBr3:Ce continuous crystals are not available yet.

Original languageEnglish
Pages (from-to)296-300
Number of pages5
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume569
Issue number2 SPEC. ISS.
DOIs
Publication statusPublished - Dec 20 2006

Fingerprint

Image sensors
Cameras
cameras
Scintillation
zinc tellurides
scintillation
cadmium tellurides
Crystals
Cadmium
Zinc
spatial resolution
crystals
detectors
Detectors
Nuclear medicine
Imaging techniques
Semiconductor detectors
nuclear medicine
energy
Lanthanum

Keywords

  • CZT
  • Gamma-ray imagers
  • Lanthanum bromide
  • Nuclear medicine
  • Scintimammography

ASJC Scopus subject areas

  • Instrumentation
  • Nuclear and High Energy Physics

Cite this

Recent advances and future perspectives of gamma imagers for scintimammography. / Pani, R.; Pellegrini, R.; Cinti, M. N.; Bennati, P.; Betti, M.; Casali, V.; Schillaci, O.; Mattioli, M.; Orsolini Cencelli, V.; Navarria, F.; Bollini, D.; Moschini, G.; Garibaldi, F.; Cusanno, F.; Iurlaro, G.; Montani, L.; Scafè, R.; de Notaristefani, F.

In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 569, No. 2 SPEC. ISS., 20.12.2006, p. 296-300.

Research output: Contribution to journalArticle

Pani, R, Pellegrini, R, Cinti, MN, Bennati, P, Betti, M, Casali, V, Schillaci, O, Mattioli, M, Orsolini Cencelli, V, Navarria, F, Bollini, D, Moschini, G, Garibaldi, F, Cusanno, F, Iurlaro, G, Montani, L, Scafè, R & de Notaristefani, F 2006, 'Recent advances and future perspectives of gamma imagers for scintimammography', Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 569, no. 2 SPEC. ISS., pp. 296-300. https://doi.org/10.1016/j.nima.2006.08.024
Pani, R. ; Pellegrini, R. ; Cinti, M. N. ; Bennati, P. ; Betti, M. ; Casali, V. ; Schillaci, O. ; Mattioli, M. ; Orsolini Cencelli, V. ; Navarria, F. ; Bollini, D. ; Moschini, G. ; Garibaldi, F. ; Cusanno, F. ; Iurlaro, G. ; Montani, L. ; Scafè, R. ; de Notaristefani, F. / Recent advances and future perspectives of gamma imagers for scintimammography. In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2006 ; Vol. 569, No. 2 SPEC. ISS. pp. 296-300.
@article{a45101a41172416cbf0f25289b12fd4b,
title = "Recent advances and future perspectives of gamma imagers for scintimammography",
abstract = "The very low sensitivity of Scintimammography for tumors under 1 cm in diameter, with current nuclear medicine cameras in use, is the major limitation in recommending this test modality for screening purposes. Recently latest generation cameras with superior imaging performances have allowed to foresee a very promising future for scintimammography. Recent technological advances obtained from our research group on the new Lanthanum scintillation crystals are now demonstrating how continuous crystals coupled to Hamamatsu Flat panel tube can overcome a number of limitations in spatial and energy resolution of scintillation arrays, carrying out 6{\%} energy resolution, and submillimeter spatial resolution values at 140 keV. In this paper three basic detection systems have been analysed: a LumaGEM 3200S gamma camera (Gamma Medica, InC., Northridge, USA), based on a cadmium-zinc-telluride (CZT) semiconductor detector with a FoV of 12.5×12.5 cm2 and with detector elements of 1.6×1.6 mm2 in size and a detector prototype consisting of a Flat panel PMT coupled to planar LaBr3:Ce scintillator. The third one is a first generation large FoV gamma camera chosen for comparison, based on PSPMTs photodetector array coupled to a NaI(Tl) scintillation crystal matrix. LaBr3:Ce and CZT cameras showed superior spatial and energy resolution than previous generation one based on NaI(Tl) scintillation array. Both CZT and LaBr gamma cameras showed similar efficiencies for the same energy window though LaBr3:Ce continuous crystal showed better imaging performance than pixellated detectors. However large area LaBr3:Ce continuous crystals are not available yet.",
keywords = "CZT, Gamma-ray imagers, Lanthanum bromide, Nuclear medicine, Scintimammography",
author = "R. Pani and R. Pellegrini and Cinti, {M. N.} and P. Bennati and M. Betti and V. Casali and O. Schillaci and M. Mattioli and {Orsolini Cencelli}, V. and F. Navarria and D. Bollini and G. Moschini and F. Garibaldi and F. Cusanno and G. Iurlaro and L. Montani and R. Scaf{\`e} and {de Notaristefani}, F.",
year = "2006",
month = "12",
day = "20",
doi = "10.1016/j.nima.2006.08.024",
language = "English",
volume = "569",
pages = "296--300",
journal = "Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment",
issn = "0168-9002",
publisher = "Elsevier",
number = "2 SPEC. ISS.",

}

TY - JOUR

T1 - Recent advances and future perspectives of gamma imagers for scintimammography

AU - Pani, R.

AU - Pellegrini, R.

AU - Cinti, M. N.

AU - Bennati, P.

AU - Betti, M.

AU - Casali, V.

AU - Schillaci, O.

AU - Mattioli, M.

AU - Orsolini Cencelli, V.

AU - Navarria, F.

AU - Bollini, D.

AU - Moschini, G.

AU - Garibaldi, F.

AU - Cusanno, F.

AU - Iurlaro, G.

AU - Montani, L.

AU - Scafè, R.

AU - de Notaristefani, F.

PY - 2006/12/20

Y1 - 2006/12/20

N2 - The very low sensitivity of Scintimammography for tumors under 1 cm in diameter, with current nuclear medicine cameras in use, is the major limitation in recommending this test modality for screening purposes. Recently latest generation cameras with superior imaging performances have allowed to foresee a very promising future for scintimammography. Recent technological advances obtained from our research group on the new Lanthanum scintillation crystals are now demonstrating how continuous crystals coupled to Hamamatsu Flat panel tube can overcome a number of limitations in spatial and energy resolution of scintillation arrays, carrying out 6% energy resolution, and submillimeter spatial resolution values at 140 keV. In this paper three basic detection systems have been analysed: a LumaGEM 3200S gamma camera (Gamma Medica, InC., Northridge, USA), based on a cadmium-zinc-telluride (CZT) semiconductor detector with a FoV of 12.5×12.5 cm2 and with detector elements of 1.6×1.6 mm2 in size and a detector prototype consisting of a Flat panel PMT coupled to planar LaBr3:Ce scintillator. The third one is a first generation large FoV gamma camera chosen for comparison, based on PSPMTs photodetector array coupled to a NaI(Tl) scintillation crystal matrix. LaBr3:Ce and CZT cameras showed superior spatial and energy resolution than previous generation one based on NaI(Tl) scintillation array. Both CZT and LaBr gamma cameras showed similar efficiencies for the same energy window though LaBr3:Ce continuous crystal showed better imaging performance than pixellated detectors. However large area LaBr3:Ce continuous crystals are not available yet.

AB - The very low sensitivity of Scintimammography for tumors under 1 cm in diameter, with current nuclear medicine cameras in use, is the major limitation in recommending this test modality for screening purposes. Recently latest generation cameras with superior imaging performances have allowed to foresee a very promising future for scintimammography. Recent technological advances obtained from our research group on the new Lanthanum scintillation crystals are now demonstrating how continuous crystals coupled to Hamamatsu Flat panel tube can overcome a number of limitations in spatial and energy resolution of scintillation arrays, carrying out 6% energy resolution, and submillimeter spatial resolution values at 140 keV. In this paper three basic detection systems have been analysed: a LumaGEM 3200S gamma camera (Gamma Medica, InC., Northridge, USA), based on a cadmium-zinc-telluride (CZT) semiconductor detector with a FoV of 12.5×12.5 cm2 and with detector elements of 1.6×1.6 mm2 in size and a detector prototype consisting of a Flat panel PMT coupled to planar LaBr3:Ce scintillator. The third one is a first generation large FoV gamma camera chosen for comparison, based on PSPMTs photodetector array coupled to a NaI(Tl) scintillation crystal matrix. LaBr3:Ce and CZT cameras showed superior spatial and energy resolution than previous generation one based on NaI(Tl) scintillation array. Both CZT and LaBr gamma cameras showed similar efficiencies for the same energy window though LaBr3:Ce continuous crystal showed better imaging performance than pixellated detectors. However large area LaBr3:Ce continuous crystals are not available yet.

KW - CZT

KW - Gamma-ray imagers

KW - Lanthanum bromide

KW - Nuclear medicine

KW - Scintimammography

UR - http://www.scopus.com/inward/record.url?scp=33751547473&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33751547473&partnerID=8YFLogxK

U2 - 10.1016/j.nima.2006.08.024

DO - 10.1016/j.nima.2006.08.024

M3 - Article

AN - SCOPUS:33751547473

VL - 569

SP - 296

EP - 300

JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

SN - 0168-9002

IS - 2 SPEC. ISS.

ER -