Stem damage during implantation of modular hip prostheses

Elena Varini, Luca Cristofolini, Marco Viceconti, Francesco Traina

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Modular hip prostheses are commonly used today. Dedicated surgical auxiliary instrumentation is indispensable. A correct coupling between the prosthesis and the instrument is necessary to avoid damage. Based on a specific commercial design (AnCAFit, Wright Cremascoli Ortho, Toulon, France), the aim of this work was to assess if stem holders of modular prostheses can induce any stem damage, and if so, to identify the features that provide minimal damage. Two different stem holder connector designs were investigated. The surface of the stem cavity was inspected to identify the damage caused by the use of the two connectors after a simulated surgical handling. Both connectors induced some damage, with significantly more damage being caused by a supposedly improved design. Scars due to contact with the stem holder were observed on the stem cavity surface. This may be sufficient to trigger increased fretting damage during service life, causing the production of fretting debris.

Original languageEnglish
Pages (from-to)564-567
Number of pages4
JournalArtificial Organs
Volume30
Issue number7
DOIs
Publication statusPublished - Jul 2006

Fingerprint

Hip prostheses
Hip Replacement Arthroplasties
Prostheses and Implants
Hip Prosthesis
France
Cicatrix
Prosthetics
Debris
Service life

Keywords

  • Modular prostheses
  • Morse taper
  • Neck modularity
  • Surface damage
  • Surgical instrumentation
  • Total hip arthroplasty

ASJC Scopus subject areas

  • Biophysics

Cite this

Stem damage during implantation of modular hip prostheses. / Varini, Elena; Cristofolini, Luca; Viceconti, Marco; Traina, Francesco.

In: Artificial Organs, Vol. 30, No. 7, 07.2006, p. 564-567.

Research output: Contribution to journalArticle

Varini, Elena ; Cristofolini, Luca ; Viceconti, Marco ; Traina, Francesco. / Stem damage during implantation of modular hip prostheses. In: Artificial Organs. 2006 ; Vol. 30, No. 7. pp. 564-567.
@article{23760dcd9ae54399865a5c6a80b2b764,
title = "Stem damage during implantation of modular hip prostheses",
abstract = "Modular hip prostheses are commonly used today. Dedicated surgical auxiliary instrumentation is indispensable. A correct coupling between the prosthesis and the instrument is necessary to avoid damage. Based on a specific commercial design (AnCAFit, Wright Cremascoli Ortho, Toulon, France), the aim of this work was to assess if stem holders of modular prostheses can induce any stem damage, and if so, to identify the features that provide minimal damage. Two different stem holder connector designs were investigated. The surface of the stem cavity was inspected to identify the damage caused by the use of the two connectors after a simulated surgical handling. Both connectors induced some damage, with significantly more damage being caused by a supposedly improved design. Scars due to contact with the stem holder were observed on the stem cavity surface. This may be sufficient to trigger increased fretting damage during service life, causing the production of fretting debris.",
keywords = "Modular prostheses, Morse taper, Neck modularity, Surface damage, Surgical instrumentation, Total hip arthroplasty",
author = "Elena Varini and Luca Cristofolini and Marco Viceconti and Francesco Traina",
year = "2006",
month = "7",
doi = "10.1111/j.1525-1594.2006.00261.x",
language = "English",
volume = "30",
pages = "564--567",
journal = "Artificial Organs",
issn = "0160-564X",
publisher = "Wiley-Blackwell",
number = "7",

}

TY - JOUR

T1 - Stem damage during implantation of modular hip prostheses

AU - Varini, Elena

AU - Cristofolini, Luca

AU - Viceconti, Marco

AU - Traina, Francesco

PY - 2006/7

Y1 - 2006/7

N2 - Modular hip prostheses are commonly used today. Dedicated surgical auxiliary instrumentation is indispensable. A correct coupling between the prosthesis and the instrument is necessary to avoid damage. Based on a specific commercial design (AnCAFit, Wright Cremascoli Ortho, Toulon, France), the aim of this work was to assess if stem holders of modular prostheses can induce any stem damage, and if so, to identify the features that provide minimal damage. Two different stem holder connector designs were investigated. The surface of the stem cavity was inspected to identify the damage caused by the use of the two connectors after a simulated surgical handling. Both connectors induced some damage, with significantly more damage being caused by a supposedly improved design. Scars due to contact with the stem holder were observed on the stem cavity surface. This may be sufficient to trigger increased fretting damage during service life, causing the production of fretting debris.

AB - Modular hip prostheses are commonly used today. Dedicated surgical auxiliary instrumentation is indispensable. A correct coupling between the prosthesis and the instrument is necessary to avoid damage. Based on a specific commercial design (AnCAFit, Wright Cremascoli Ortho, Toulon, France), the aim of this work was to assess if stem holders of modular prostheses can induce any stem damage, and if so, to identify the features that provide minimal damage. Two different stem holder connector designs were investigated. The surface of the stem cavity was inspected to identify the damage caused by the use of the two connectors after a simulated surgical handling. Both connectors induced some damage, with significantly more damage being caused by a supposedly improved design. Scars due to contact with the stem holder were observed on the stem cavity surface. This may be sufficient to trigger increased fretting damage during service life, causing the production of fretting debris.

KW - Modular prostheses

KW - Morse taper

KW - Neck modularity

KW - Surface damage

KW - Surgical instrumentation

KW - Total hip arthroplasty

UR - http://www.scopus.com/inward/record.url?scp=33745505550&partnerID=8YFLogxK