N2 - The Zenith Dissection Endovascular System is a device designed for treatment of aortic type B dissection utilizing the Provisional ExTension to Induce COmplete ATtachment (PETTICOAT) technique. Due to the stent design (low radial force and lack of columnar support), significant risk of stent misalignment exists, which we have encountered in four out of 25 patients treated since 2005. Misalignment may result from excessive manipulation of the delivery system at the time of implantation or during catheter manipulation during adjunctive or secondary procedures. The manufacturer has modified the design of this device in order to prevent misalignment, although no serious clinical consequences of this misalignment have been reported with a mean follow-up of 50 months. Care with catheter manipulation after device deployment and accurate review of postoperative imaging are still warranted.
AB - The Zenith Dissection Endovascular System is a device designed for treatment of aortic type B dissection utilizing the Provisional ExTension to Induce COmplete ATtachment (PETTICOAT) technique. Due to the stent design (low radial force and lack of columnar support), significant risk of stent misalignment exists, which we have encountered in four out of 25 patients treated since 2005. Misalignment may result from excessive manipulation of the delivery system at the time of implantation or during catheter manipulation during adjunctive or secondary procedures. The manufacturer has modified the design of this device in order to prevent misalignment, although no serious clinical consequences of this misalignment have been reported with a mean follow-up of 50 months. Care with catheter manipulation after device deployment and accurate review of postoperative imaging are still warranted.
UR - http://www.scopus.com/inward/record.url?scp=84872686904&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84872686904&partnerID=8YFLogxK
U2 - 10.1016/j.jvs.2012.07.004
DO - 10.1016/j.jvs.2012.07.004
M3 - Article
C2 - 22975337
AN - SCOPUS:84872686904
VL - 57
SP - 515
EP - 517
JO - Journal of Vascular Surgery
JF - Journal of Vascular Surgery
SN - 0741-5214
IS - 2
ER -