Valutazione del "recoil" acuto dello stent di Palmaz-Schatz mediante ecografia. Descrizione di un caso.

Translated title of the contribution: The evaluation of the acute recoil of the Palmaz-Schatz stent using echography. A case report

L. Maiello, A. Itoh, A. Colombo

Research output: Contribution to journalArticlepeer-review

Abstract

Intravascular ultrasound, performed using a 0.018 in guide wire with a 30 MHz ultrasound transducer, allows to study the acute recoil of Palmaz-Schatz stent. This case showed that the acute recoil of the Palmaz-Schatz stent occurs immediately after stent implantation. The analysis of diameter and lumen cross-sectional area at the stent site was performed at different levels of dilatation. A 20% reduction of cross-sectional area was found immediately after stent expansion and balloon deflation.

Translated title of the contributionThe evaluation of the acute recoil of the Palmaz-Schatz stent using echography. A case report
Original languageItalian
Pages (from-to)559-562
Number of pages4
JournalCardiologia
Volume41
Issue number6
Publication statusPublished - Jun 1996