Influenza sul tracciato elettroretinografico di elettrodi attivi di diversa concezione

Translated title of the contribution: The influence of different active electrodes on electroretinographic tracing

A. Valentini, R. Ceccuzzi, E. Poppi, P. Piccinini, F. Sandolo, G. Furiosi, C. Gelmi

Research output: Contribution to journalArticle

Abstract

In this work we have considered the electroretinographic tracing, achieved with the same stimulating parameters making use of four different types of electrodes: corneale, gold-foil, palpebral and DTL. We have estimated the a and b waves amplitudes and we considered the mean±ds for each tracing executed with the same electrodes. The corneal electrode let we record the biggest amplitude wave. The DTL, gold-foil and palpebral electrode tracing gave us a lower amplitude. To obtain comparable clinical results we calculated conversion factors for the tracing amplitudes recorded with the different types of active electrodes.

Translated title of the contributionThe influence of different active electrodes on electroretinographic tracing
Original languageItalian
Pages (from-to)145-150
Number of pages6
JournalAnnali di Ottalmologia e Clinica Oculistica
Volume124
Issue number5-6
Publication statusPublished - 1998

ASJC Scopus subject areas

  • Ophthalmology

Fingerprint Dive into the research topics of 'The influence of different active electrodes on electroretinographic tracing'. Together they form a unique fingerprint.