The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2: Wear particle characteristics with isolation protocols

S. Affatato, G. Bersaglia, D. Emiliani, I. Foltran, P. Taddei, M. Reggiani, P. Ferrieri, A. Toni

Research output: Contribution to journalArticle

Abstract

Ultra-high molecular-weight polyethylene (UHMWPE) has been used in total joint replacement for the last three decades. Despite the recent advancements in prosthesis design, the wear of UHMWPE remains a serious clinical problem; the release of wear debris may induce osteolysis and implant loosening. Controlling the quality of the polyethylene is essential to improve its wear resistance and any potential adverse effect caused by processing, manufacturing or sterilisation should be avoided. To evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide (EtO)-treatment) and third-body particles, gamma- and EtO-sterilised UHMWPE acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5million cycles in bovine calf serum in the presence of third-body polymethylmethacrylate (PMMA) particles. A method not requiring ultra-centrifugation has been proposed for the isolation of polyethylene wear debris from the serum lubricant. SEM analysis allowed debris shape and morphology to be determined, and the wear mechanism operating in this study to be hypothesised. The morphological features of the wear debris were in agreement with clinical findings, enabling the hip simulator function to be validated. Micro-Raman spectroscopy coupled to PLS analysis showed that the mechanical friction during in vitro tests induced significant crystallinity changes in all the cups. The most significant changes were observed for the EtO-sterilised cups, which showed the highest weight loss.

Original languageEnglish
Pages (from-to)4045-4055
Number of pages11
JournalBiomaterials
Volume24
Issue number22
DOIs
Publication statusPublished - Oct 2003

Fingerprint

Ethylene Oxide
Ultrahigh molecular weight polyethylenes
Ethylene
Debris
Simulators
Wear of materials
Polyethylene
Oxides
Replacement Arthroplasties
Prosthesis Design
Lubricants
Osteolysis
Raman Spectrum Analysis
Friction
Hip Joint
Polymethyl Methacrylate
Polyethylenes
Thigh
Serum
Centrifugation

Keywords

  • Micro-Raman spectroscopy
  • Polyethylene debris
  • SEM analysis
  • Sterilisation method

ASJC Scopus subject areas

  • Biotechnology
  • Bioengineering
  • Biomedical Engineering

Cite this

The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2 : Wear particle characteristics with isolation protocols. / Affatato, S.; Bersaglia, G.; Emiliani, D.; Foltran, I.; Taddei, P.; Reggiani, M.; Ferrieri, P.; Toni, A.

In: Biomaterials, Vol. 24, No. 22, 10.2003, p. 4045-4055.

Research output: Contribution to journalArticle

Affatato, S. ; Bersaglia, G. ; Emiliani, D. ; Foltran, I. ; Taddei, P. ; Reggiani, M. ; Ferrieri, P. ; Toni, A. / The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2 : Wear particle characteristics with isolation protocols. In: Biomaterials. 2003 ; Vol. 24, No. 22. pp. 4045-4055.
@article{718c8352b84849da8f294c1a209200e3,
title = "The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2: Wear particle characteristics with isolation protocols",
abstract = "Ultra-high molecular-weight polyethylene (UHMWPE) has been used in total joint replacement for the last three decades. Despite the recent advancements in prosthesis design, the wear of UHMWPE remains a serious clinical problem; the release of wear debris may induce osteolysis and implant loosening. Controlling the quality of the polyethylene is essential to improve its wear resistance and any potential adverse effect caused by processing, manufacturing or sterilisation should be avoided. To evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide (EtO)-treatment) and third-body particles, gamma- and EtO-sterilised UHMWPE acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5million cycles in bovine calf serum in the presence of third-body polymethylmethacrylate (PMMA) particles. A method not requiring ultra-centrifugation has been proposed for the isolation of polyethylene wear debris from the serum lubricant. SEM analysis allowed debris shape and morphology to be determined, and the wear mechanism operating in this study to be hypothesised. The morphological features of the wear debris were in agreement with clinical findings, enabling the hip simulator function to be validated. Micro-Raman spectroscopy coupled to PLS analysis showed that the mechanical friction during in vitro tests induced significant crystallinity changes in all the cups. The most significant changes were observed for the EtO-sterilised cups, which showed the highest weight loss.",
keywords = "Micro-Raman spectroscopy, Polyethylene debris, SEM analysis, Sterilisation method",
author = "S. Affatato and G. Bersaglia and D. Emiliani and I. Foltran and P. Taddei and M. Reggiani and P. Ferrieri and A. Toni",
year = "2003",
month = "10",
doi = "10.1016/S0142-9612(03)00264-3",
language = "English",
volume = "24",
pages = "4045--4055",
journal = "Biomaterials",
issn = "0142-9612",
publisher = "Elsevier BV",
number = "22",

}

TY - JOUR

T1 - The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2

T2 - Wear particle characteristics with isolation protocols

AU - Affatato, S.

AU - Bersaglia, G.

AU - Emiliani, D.

AU - Foltran, I.

AU - Taddei, P.

AU - Reggiani, M.

AU - Ferrieri, P.

AU - Toni, A.

PY - 2003/10

Y1 - 2003/10

N2 - Ultra-high molecular-weight polyethylene (UHMWPE) has been used in total joint replacement for the last three decades. Despite the recent advancements in prosthesis design, the wear of UHMWPE remains a serious clinical problem; the release of wear debris may induce osteolysis and implant loosening. Controlling the quality of the polyethylene is essential to improve its wear resistance and any potential adverse effect caused by processing, manufacturing or sterilisation should be avoided. To evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide (EtO)-treatment) and third-body particles, gamma- and EtO-sterilised UHMWPE acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5million cycles in bovine calf serum in the presence of third-body polymethylmethacrylate (PMMA) particles. A method not requiring ultra-centrifugation has been proposed for the isolation of polyethylene wear debris from the serum lubricant. SEM analysis allowed debris shape and morphology to be determined, and the wear mechanism operating in this study to be hypothesised. The morphological features of the wear debris were in agreement with clinical findings, enabling the hip simulator function to be validated. Micro-Raman spectroscopy coupled to PLS analysis showed that the mechanical friction during in vitro tests induced significant crystallinity changes in all the cups. The most significant changes were observed for the EtO-sterilised cups, which showed the highest weight loss.

AB - Ultra-high molecular-weight polyethylene (UHMWPE) has been used in total joint replacement for the last three decades. Despite the recent advancements in prosthesis design, the wear of UHMWPE remains a serious clinical problem; the release of wear debris may induce osteolysis and implant loosening. Controlling the quality of the polyethylene is essential to improve its wear resistance and any potential adverse effect caused by processing, manufacturing or sterilisation should be avoided. To evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide (EtO)-treatment) and third-body particles, gamma- and EtO-sterilised UHMWPE acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5million cycles in bovine calf serum in the presence of third-body polymethylmethacrylate (PMMA) particles. A method not requiring ultra-centrifugation has been proposed for the isolation of polyethylene wear debris from the serum lubricant. SEM analysis allowed debris shape and morphology to be determined, and the wear mechanism operating in this study to be hypothesised. The morphological features of the wear debris were in agreement with clinical findings, enabling the hip simulator function to be validated. Micro-Raman spectroscopy coupled to PLS analysis showed that the mechanical friction during in vitro tests induced significant crystallinity changes in all the cups. The most significant changes were observed for the EtO-sterilised cups, which showed the highest weight loss.

KW - Micro-Raman spectroscopy

KW - Polyethylene debris

KW - SEM analysis

KW - Sterilisation method

UR - http://www.scopus.com/inward/record.url?scp=0037813128&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037813128&partnerID=8YFLogxK

U2 - 10.1016/S0142-9612(03)00264-3

DO - 10.1016/S0142-9612(03)00264-3

M3 - Article

C2 - 12834600

AN - SCOPUS:0037813128

VL - 24

SP - 4045

EP - 4055

JO - Biomaterials

JF - Biomaterials

SN - 0142-9612

IS - 22

ER -