The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: Role of the third-body wear process

S. Affatato, G. Bersaglia, I. Foltran, P. Taddei, G. Fini, A. Toni

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

Due to its excellent combination of properties, ultra-high-molecular-weight-polyethylene has been used for the last 30 years in the replacement of damaged articulating cartilage for total joint replacement surgery. However, in some cases, wear, failure and delamination have been observed. Polyethylene performance may be affected by oxidation during consolidation of the resin, sterilisation of the finished specimens and post-irradiation storage. In order to evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide(EtO)-treatment) and third-body particles on the ultra-high-molecular-weight-polyethylene wear behaviour, gamma- and EtO-sterilised ultra-high-molecular-weight-polyethylene acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body PMMA particles. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a significantly lower wear rate than those EtO-sterilised. Moreover, significant differences were found for each type of sterilisation between the gravimetric wear trends obtained until 2.5 million cycles in the presence and in the absence of PMMA particles.

Original languageEnglish
Pages (from-to)4839-4846
Number of pages8
JournalBiomaterials
Volume23
Issue number24
DOIs
Publication statusPublished - Dec 2002

Fingerprint

Ethylene Oxide
Ultrahigh molecular weight polyethylenes
Ethylene
Simulators
Wear of materials
Polymethyl Methacrylate
Oxides
Replacement Arthroplasties
Hip Joint
Irradiation
Polyethylene
Thigh
Cartilage
Weight Loss
Delamination
Consolidation
Surgery
Polyethylenes
Resins
Serum

Keywords

  • Debris
  • Morphology
  • SEM analysis
  • Sterilisation
  • Third-body wear process

ASJC Scopus subject areas

  • Biotechnology
  • Bioengineering
  • Biomedical Engineering

Cite this

The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1 : Role of the third-body wear process. / Affatato, S.; Bersaglia, G.; Foltran, I.; Taddei, P.; Fini, G.; Toni, A.

In: Biomaterials, Vol. 23, No. 24, 12.2002, p. 4839-4846.

Research output: Contribution to journalArticle

@article{e1ba53378d5c48da809ebe75624b2b4b,
title = "The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: Role of the third-body wear process",
abstract = "Due to its excellent combination of properties, ultra-high-molecular-weight-polyethylene has been used for the last 30 years in the replacement of damaged articulating cartilage for total joint replacement surgery. However, in some cases, wear, failure and delamination have been observed. Polyethylene performance may be affected by oxidation during consolidation of the resin, sterilisation of the finished specimens and post-irradiation storage. In order to evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide(EtO)-treatment) and third-body particles on the ultra-high-molecular-weight-polyethylene wear behaviour, gamma- and EtO-sterilised ultra-high-molecular-weight-polyethylene acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body PMMA particles. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a significantly lower wear rate than those EtO-sterilised. Moreover, significant differences were found for each type of sterilisation between the gravimetric wear trends obtained until 2.5 million cycles in the presence and in the absence of PMMA particles.",
keywords = "Debris, Morphology, SEM analysis, Sterilisation, Third-body wear process",
author = "S. Affatato and G. Bersaglia and I. Foltran and P. Taddei and G. Fini and A. Toni",
year = "2002",
month = "12",
doi = "10.1016/S0142-9612(02)00238-7",
language = "English",
volume = "23",
pages = "4839--4846",
journal = "Biomaterials",
issn = "0142-9612",
publisher = "Elsevier BV",
number = "24",

}

TY - JOUR

T1 - The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1

T2 - Role of the third-body wear process

AU - Affatato, S.

AU - Bersaglia, G.

AU - Foltran, I.

AU - Taddei, P.

AU - Fini, G.

AU - Toni, A.

PY - 2002/12

Y1 - 2002/12

N2 - Due to its excellent combination of properties, ultra-high-molecular-weight-polyethylene has been used for the last 30 years in the replacement of damaged articulating cartilage for total joint replacement surgery. However, in some cases, wear, failure and delamination have been observed. Polyethylene performance may be affected by oxidation during consolidation of the resin, sterilisation of the finished specimens and post-irradiation storage. In order to evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide(EtO)-treatment) and third-body particles on the ultra-high-molecular-weight-polyethylene wear behaviour, gamma- and EtO-sterilised ultra-high-molecular-weight-polyethylene acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body PMMA particles. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a significantly lower wear rate than those EtO-sterilised. Moreover, significant differences were found for each type of sterilisation between the gravimetric wear trends obtained until 2.5 million cycles in the presence and in the absence of PMMA particles.

AB - Due to its excellent combination of properties, ultra-high-molecular-weight-polyethylene has been used for the last 30 years in the replacement of damaged articulating cartilage for total joint replacement surgery. However, in some cases, wear, failure and delamination have been observed. Polyethylene performance may be affected by oxidation during consolidation of the resin, sterilisation of the finished specimens and post-irradiation storage. In order to evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide(EtO)-treatment) and third-body particles on the ultra-high-molecular-weight-polyethylene wear behaviour, gamma- and EtO-sterilised ultra-high-molecular-weight-polyethylene acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body PMMA particles. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a significantly lower wear rate than those EtO-sterilised. Moreover, significant differences were found for each type of sterilisation between the gravimetric wear trends obtained until 2.5 million cycles in the presence and in the absence of PMMA particles.

KW - Debris

KW - Morphology

KW - SEM analysis

KW - Sterilisation

KW - Third-body wear process

UR - http://www.scopus.com/inward/record.url?scp=0036888509&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036888509&partnerID=8YFLogxK

U2 - 10.1016/S0142-9612(02)00238-7

DO - 10.1016/S0142-9612(02)00238-7

M3 - Article

C2 - 12361624

AN - SCOPUS:0036888509

VL - 23

SP - 4839

EP - 4846

JO - Biomaterials

JF - Biomaterials

SN - 0142-9612

IS - 24

ER -