The wall correction factor for a spherical ionization chamber used in brachytherapy source calibration

A. Piermattei, L. Azario, A. Fidanzio, P. Viola, C. Dell'Omo, L. Iadanza, V. Fusco, J. I. Lagares, R. Capote

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

The effect of wall chamber attenuation and scattering is one of the most important corrections that must be detennined when the linear interpolation method between two calibration factors of an ionization chamber is used. For spherical ionization chambers the corresponding correction factors Aw have to be determined by a non-linear trend of the response as a function of the wall thickness. The Monte Carlo and experimental data here reported show that the Aw factors obtained for an Exradin A4 chamber, used in the brachytherapy source calibration, in terms of reference air kerma rate, are up to 1.2% greater than the values obtained by the linear extrapolation method for the studied beam qualities. Using the Aw factors derived from Monte Carlo calculations, the accuracy of the calibration factor NK,Ir for the Exradin A4, obtained by the interpolation between two calibration factors, improves about 0.6%. The discrepancy between the new calculated factor and that obtained using the complete calibration curve of the ion-chamber and the 192Ir spectrum is only 0.1%.

Original languageEnglish
Pages (from-to)4091-4103
Number of pages13
JournalPhysics in Medicine and Biology
Volume48
Issue number24
DOIs
Publication statusPublished - Dec 21 2003

Fingerprint

Ionization chambers
Brachytherapy
ionization chambers
Calibration
Interpolation
interpolation
Beam quality
chambers
Extrapolation
Air
Scattering
Ions
extrapolation
attenuation
trends
air
curves
scattering

ASJC Scopus subject areas

  • Biomedical Engineering
  • Physics and Astronomy (miscellaneous)
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

The wall correction factor for a spherical ionization chamber used in brachytherapy source calibration. / Piermattei, A.; Azario, L.; Fidanzio, A.; Viola, P.; Dell'Omo, C.; Iadanza, L.; Fusco, V.; Lagares, J. I.; Capote, R.

In: Physics in Medicine and Biology, Vol. 48, No. 24, 21.12.2003, p. 4091-4103.

Research output: Contribution to journalArticle

Piermattei, A. ; Azario, L. ; Fidanzio, A. ; Viola, P. ; Dell'Omo, C. ; Iadanza, L. ; Fusco, V. ; Lagares, J. I. ; Capote, R. / The wall correction factor for a spherical ionization chamber used in brachytherapy source calibration. In: Physics in Medicine and Biology. 2003 ; Vol. 48, No. 24. pp. 4091-4103.
@article{d14907d9d54a4ab0b5ad421daec285f8,
title = "The wall correction factor for a spherical ionization chamber used in brachytherapy source calibration",
abstract = "The effect of wall chamber attenuation and scattering is one of the most important corrections that must be detennined when the linear interpolation method between two calibration factors of an ionization chamber is used. For spherical ionization chambers the corresponding correction factors Aw have to be determined by a non-linear trend of the response as a function of the wall thickness. The Monte Carlo and experimental data here reported show that the Aw factors obtained for an Exradin A4 chamber, used in the brachytherapy source calibration, in terms of reference air kerma rate, are up to 1.2{\%} greater than the values obtained by the linear extrapolation method for the studied beam qualities. Using the Aw factors derived from Monte Carlo calculations, the accuracy of the calibration factor NK,Ir for the Exradin A4, obtained by the interpolation between two calibration factors, improves about 0.6{\%}. The discrepancy between the new calculated factor and that obtained using the complete calibration curve of the ion-chamber and the 192Ir spectrum is only 0.1{\%}.",
author = "A. Piermattei and L. Azario and A. Fidanzio and P. Viola and C. Dell'Omo and L. Iadanza and V. Fusco and Lagares, {J. I.} and R. Capote",
year = "2003",
month = "12",
day = "21",
doi = "10.1088/0031-9155/48/24/007",
language = "English",
volume = "48",
pages = "4091--4103",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "24",

}

TY - JOUR

T1 - The wall correction factor for a spherical ionization chamber used in brachytherapy source calibration

AU - Piermattei, A.

AU - Azario, L.

AU - Fidanzio, A.

AU - Viola, P.

AU - Dell'Omo, C.

AU - Iadanza, L.

AU - Fusco, V.

AU - Lagares, J. I.

AU - Capote, R.

PY - 2003/12/21

Y1 - 2003/12/21

N2 - The effect of wall chamber attenuation and scattering is one of the most important corrections that must be detennined when the linear interpolation method between two calibration factors of an ionization chamber is used. For spherical ionization chambers the corresponding correction factors Aw have to be determined by a non-linear trend of the response as a function of the wall thickness. The Monte Carlo and experimental data here reported show that the Aw factors obtained for an Exradin A4 chamber, used in the brachytherapy source calibration, in terms of reference air kerma rate, are up to 1.2% greater than the values obtained by the linear extrapolation method for the studied beam qualities. Using the Aw factors derived from Monte Carlo calculations, the accuracy of the calibration factor NK,Ir for the Exradin A4, obtained by the interpolation between two calibration factors, improves about 0.6%. The discrepancy between the new calculated factor and that obtained using the complete calibration curve of the ion-chamber and the 192Ir spectrum is only 0.1%.

AB - The effect of wall chamber attenuation and scattering is one of the most important corrections that must be detennined when the linear interpolation method between two calibration factors of an ionization chamber is used. For spherical ionization chambers the corresponding correction factors Aw have to be determined by a non-linear trend of the response as a function of the wall thickness. The Monte Carlo and experimental data here reported show that the Aw factors obtained for an Exradin A4 chamber, used in the brachytherapy source calibration, in terms of reference air kerma rate, are up to 1.2% greater than the values obtained by the linear extrapolation method for the studied beam qualities. Using the Aw factors derived from Monte Carlo calculations, the accuracy of the calibration factor NK,Ir for the Exradin A4, obtained by the interpolation between two calibration factors, improves about 0.6%. The discrepancy between the new calculated factor and that obtained using the complete calibration curve of the ion-chamber and the 192Ir spectrum is only 0.1%.

UR - http://www.scopus.com/inward/record.url?scp=0346336961&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0346336961&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/48/24/007

DO - 10.1088/0031-9155/48/24/007

M3 - Article

VL - 48

SP - 4091

EP - 4103

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 24

ER -