Thermodynamics of vitreoretinal surgery

Mario R. Romano, Jose L. Vallejo-Garcia, Vito Romano, Martina Angi, Paolo Vinciguerra, Ciro Costagliola

Research output: Contribution to journalArticle

Abstract

Purpose: To report in vivo the temperature variations occurring continuously inside the human eye during vitreoretinal surgery. Methods: Intraocular temperature was monitored during the entire surgical time in 14 eyes of 14 patients undergoing pars plana vitrectomy. A custom made 23-gauge thermoprobe was inserted through a sclerotomy at 3.5 mm from the limbus. Temperature in the anterior chamber and on the retinal surface was measured at one time point, whilst the temperature in the vitreous cavity was monitored throughout the entire surgery. Time points of particular interest in the vitreous cavity were: at baseline (inflow off-outflow off), at the beginning of the vitrectomy (inflow off-outflow on), at the end of vitrectomy (inflow on-outflow on), during the epiretinal procedures (inflow on-outflow off), under air infusion, under oil and under endolaser treatment ongoing. Results: The mean temperature in the anterior chamber and on the nasal retinal side was 23.6°C (SD 1.8)and 32.3°C (SD 1.1), respectively. The mean vitreous temperature at baseline, before opening the infusion, was 33.6°C (SD 1.4). The mean temperature at the beginning and the end of the vitrectomy was 26.8°C (SD 1.0) and 24.8°C (SD 0.8), respectively. During the epiretinal procedures, the mean temperature was 27.4°C (SD 0.7). Under air infusion, the mean temperature increased to 29.2°C (SD 1.2). The fluctuations in temperature were statistically different between the time points during the vitrectomy (p <0.001). Conclusion: Vitreoretinal surgery induces measurable changes in temperature in the human eye, with significant fluctuations. In particular, the temperature in the vitreous cavity decreases when both the inflow and the outflow fluid lines are open, whilst it increases under air infusion.

Original languageEnglish
Pages (from-to)371-374
Number of pages4
JournalCurrent Eye Research
Volume38
Issue number3
DOIs
Publication statusPublished - Mar 2013

Fingerprint

Vitreoretinal Surgery
Thermodynamics
Temperature
Vitrectomy
Air
Anterior Chamber
Temazepam
Operative Time
Nose

Keywords

  • Intraocular temperature
  • Perfluorocarbon liquid vapor tension
  • Tempearature in vitreous cavity
  • Thermodynamics of vitreoretinal surgery

ASJC Scopus subject areas

  • Ophthalmology
  • Sensory Systems
  • Cellular and Molecular Neuroscience

Cite this

Thermodynamics of vitreoretinal surgery. / Romano, Mario R.; Vallejo-Garcia, Jose L.; Romano, Vito; Angi, Martina; Vinciguerra, Paolo; Costagliola, Ciro.

In: Current Eye Research, Vol. 38, No. 3, 03.2013, p. 371-374.

Research output: Contribution to journalArticle

Romano, Mario R. ; Vallejo-Garcia, Jose L. ; Romano, Vito ; Angi, Martina ; Vinciguerra, Paolo ; Costagliola, Ciro. / Thermodynamics of vitreoretinal surgery. In: Current Eye Research. 2013 ; Vol. 38, No. 3. pp. 371-374.
@article{ec2c77c4e8b54d89b15ec6fb2e32983b,
title = "Thermodynamics of vitreoretinal surgery",
abstract = "Purpose: To report in vivo the temperature variations occurring continuously inside the human eye during vitreoretinal surgery. Methods: Intraocular temperature was monitored during the entire surgical time in 14 eyes of 14 patients undergoing pars plana vitrectomy. A custom made 23-gauge thermoprobe was inserted through a sclerotomy at 3.5 mm from the limbus. Temperature in the anterior chamber and on the retinal surface was measured at one time point, whilst the temperature in the vitreous cavity was monitored throughout the entire surgery. Time points of particular interest in the vitreous cavity were: at baseline (inflow off-outflow off), at the beginning of the vitrectomy (inflow off-outflow on), at the end of vitrectomy (inflow on-outflow on), during the epiretinal procedures (inflow on-outflow off), under air infusion, under oil and under endolaser treatment ongoing. Results: The mean temperature in the anterior chamber and on the nasal retinal side was 23.6°C (SD 1.8)and 32.3°C (SD 1.1), respectively. The mean vitreous temperature at baseline, before opening the infusion, was 33.6°C (SD 1.4). The mean temperature at the beginning and the end of the vitrectomy was 26.8°C (SD 1.0) and 24.8°C (SD 0.8), respectively. During the epiretinal procedures, the mean temperature was 27.4°C (SD 0.7). Under air infusion, the mean temperature increased to 29.2°C (SD 1.2). The fluctuations in temperature were statistically different between the time points during the vitrectomy (p <0.001). Conclusion: Vitreoretinal surgery induces measurable changes in temperature in the human eye, with significant fluctuations. In particular, the temperature in the vitreous cavity decreases when both the inflow and the outflow fluid lines are open, whilst it increases under air infusion.",
keywords = "Intraocular temperature, Perfluorocarbon liquid vapor tension, Tempearature in vitreous cavity, Thermodynamics of vitreoretinal surgery",
author = "Romano, {Mario R.} and Vallejo-Garcia, {Jose L.} and Vito Romano and Martina Angi and Paolo Vinciguerra and Ciro Costagliola",
year = "2013",
month = "3",
doi = "10.3109/02713683.2012.745160",
language = "English",
volume = "38",
pages = "371--374",
journal = "Current Eye Research",
issn = "0271-3683",
publisher = "Informa Healthcare",
number = "3",

}

TY - JOUR

T1 - Thermodynamics of vitreoretinal surgery

AU - Romano, Mario R.

AU - Vallejo-Garcia, Jose L.

AU - Romano, Vito

AU - Angi, Martina

AU - Vinciguerra, Paolo

AU - Costagliola, Ciro

PY - 2013/3

Y1 - 2013/3

N2 - Purpose: To report in vivo the temperature variations occurring continuously inside the human eye during vitreoretinal surgery. Methods: Intraocular temperature was monitored during the entire surgical time in 14 eyes of 14 patients undergoing pars plana vitrectomy. A custom made 23-gauge thermoprobe was inserted through a sclerotomy at 3.5 mm from the limbus. Temperature in the anterior chamber and on the retinal surface was measured at one time point, whilst the temperature in the vitreous cavity was monitored throughout the entire surgery. Time points of particular interest in the vitreous cavity were: at baseline (inflow off-outflow off), at the beginning of the vitrectomy (inflow off-outflow on), at the end of vitrectomy (inflow on-outflow on), during the epiretinal procedures (inflow on-outflow off), under air infusion, under oil and under endolaser treatment ongoing. Results: The mean temperature in the anterior chamber and on the nasal retinal side was 23.6°C (SD 1.8)and 32.3°C (SD 1.1), respectively. The mean vitreous temperature at baseline, before opening the infusion, was 33.6°C (SD 1.4). The mean temperature at the beginning and the end of the vitrectomy was 26.8°C (SD 1.0) and 24.8°C (SD 0.8), respectively. During the epiretinal procedures, the mean temperature was 27.4°C (SD 0.7). Under air infusion, the mean temperature increased to 29.2°C (SD 1.2). The fluctuations in temperature were statistically different between the time points during the vitrectomy (p <0.001). Conclusion: Vitreoretinal surgery induces measurable changes in temperature in the human eye, with significant fluctuations. In particular, the temperature in the vitreous cavity decreases when both the inflow and the outflow fluid lines are open, whilst it increases under air infusion.

AB - Purpose: To report in vivo the temperature variations occurring continuously inside the human eye during vitreoretinal surgery. Methods: Intraocular temperature was monitored during the entire surgical time in 14 eyes of 14 patients undergoing pars plana vitrectomy. A custom made 23-gauge thermoprobe was inserted through a sclerotomy at 3.5 mm from the limbus. Temperature in the anterior chamber and on the retinal surface was measured at one time point, whilst the temperature in the vitreous cavity was monitored throughout the entire surgery. Time points of particular interest in the vitreous cavity were: at baseline (inflow off-outflow off), at the beginning of the vitrectomy (inflow off-outflow on), at the end of vitrectomy (inflow on-outflow on), during the epiretinal procedures (inflow on-outflow off), under air infusion, under oil and under endolaser treatment ongoing. Results: The mean temperature in the anterior chamber and on the nasal retinal side was 23.6°C (SD 1.8)and 32.3°C (SD 1.1), respectively. The mean vitreous temperature at baseline, before opening the infusion, was 33.6°C (SD 1.4). The mean temperature at the beginning and the end of the vitrectomy was 26.8°C (SD 1.0) and 24.8°C (SD 0.8), respectively. During the epiretinal procedures, the mean temperature was 27.4°C (SD 0.7). Under air infusion, the mean temperature increased to 29.2°C (SD 1.2). The fluctuations in temperature were statistically different between the time points during the vitrectomy (p <0.001). Conclusion: Vitreoretinal surgery induces measurable changes in temperature in the human eye, with significant fluctuations. In particular, the temperature in the vitreous cavity decreases when both the inflow and the outflow fluid lines are open, whilst it increases under air infusion.

KW - Intraocular temperature

KW - Perfluorocarbon liquid vapor tension

KW - Tempearature in vitreous cavity

KW - Thermodynamics of vitreoretinal surgery

UR - http://www.scopus.com/inward/record.url?scp=84874488331&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84874488331&partnerID=8YFLogxK

U2 - 10.3109/02713683.2012.745160

DO - 10.3109/02713683.2012.745160

M3 - Article

C2 - 23230955

AN - SCOPUS:84874488331

VL - 38

SP - 371

EP - 374

JO - Current Eye Research

JF - Current Eye Research

SN - 0271-3683

IS - 3

ER -