Total respiratory support with tidal flow extracorporeal circulation in adult sheep

Srinivas Kolla, Stefania Crotti, W. Anthony Lee, Matthew J. Gargulinski, Thomas Lewandowski, David Bach, Ronald B. Hirschl, Robert H. Bartlett

Research output: Contribution to journalArticle

Abstract

A novel pressure gated tidal flow extracorporeal circulation (TF ECC) device was developed, and it was hypothesized that it could provide total respiratory support in apneic adult sheep without adverse hemodynamic or cardiac effects. The circuit consisted of a single lumen cannula, computer driven tubing occluders gated by circuit pressure, a nonocclusive peristaltic blood pump, a spiral coiled membrane lung, and a heat exchanger. Six paralyzed, anesthetized adult sheep were instrumented and TF ECC was instituted via cannulation of the right atrium. Total respiratory support was provided by the circuit during an apneic period of 6 hours. Echocardiography was performed with the animal instrumented (baseline) and after 2 hours of TF ECC. Circuit blood tidal volume was 172.6 ± 18.0 cc, resulting in a TF ECC flow of 71.1 ± 10.1 cc/kg/min. At the end of the study period, P(a)Co2 was 35.5 ± 7.6 mmHg, P(a)O2 was 91.2 ± 30.6 mmHg, and pulmonary artery oxygen saturation (S(PA)O2)) was 95 ± 5%. Hemodynamic stability was maintained with no significant differences at baseline and after 6 hours in mean arterial pressure, mean pulmonary artery pressure, or heart rate noted. Echocardiographic evaluation showed preserved fractional shortening of the left ventricular (LV) septal-lateral dimension (baseline 32.4 ± 11.4 %; 2 hours 34.8 ± 8.4 %). This study demonstrates TF ECC provides total respiratory support without adverse hemodynamic effects, and preserved LV function.

Original languageEnglish
JournalASAIO Journal
Volume43
Issue number5
Publication statusPublished - Sep 1997

Fingerprint

Extracorporeal Circulation
Sheep
Hemodynamics
Networks (circuits)
Blood
Pressure
Pulmonary Artery
Echocardiography
Tubing
Heat exchangers
Tidal Volume
Animals
Blood Volume
Heart Atria
Left Ventricular Function
Catheterization
Pumps
Oxygen
Membranes
Arterial Pressure

ASJC Scopus subject areas

  • Biophysics
  • Bioengineering

Cite this

Kolla, S., Crotti, S., Lee, W. A., Gargulinski, M. J., Lewandowski, T., Bach, D., ... Bartlett, R. H. (1997). Total respiratory support with tidal flow extracorporeal circulation in adult sheep. ASAIO Journal, 43(5).

Total respiratory support with tidal flow extracorporeal circulation in adult sheep. / Kolla, Srinivas; Crotti, Stefania; Lee, W. Anthony; Gargulinski, Matthew J.; Lewandowski, Thomas; Bach, David; Hirschl, Ronald B.; Bartlett, Robert H.

In: ASAIO Journal, Vol. 43, No. 5, 09.1997.

Research output: Contribution to journalArticle

Kolla, S, Crotti, S, Lee, WA, Gargulinski, MJ, Lewandowski, T, Bach, D, Hirschl, RB & Bartlett, RH 1997, 'Total respiratory support with tidal flow extracorporeal circulation in adult sheep', ASAIO Journal, vol. 43, no. 5.
Kolla S, Crotti S, Lee WA, Gargulinski MJ, Lewandowski T, Bach D et al. Total respiratory support with tidal flow extracorporeal circulation in adult sheep. ASAIO Journal. 1997 Sep;43(5).
Kolla, Srinivas ; Crotti, Stefania ; Lee, W. Anthony ; Gargulinski, Matthew J. ; Lewandowski, Thomas ; Bach, David ; Hirschl, Ronald B. ; Bartlett, Robert H. / Total respiratory support with tidal flow extracorporeal circulation in adult sheep. In: ASAIO Journal. 1997 ; Vol. 43, No. 5.
@article{0fa17d8279e84ba7adef0d1433fe0223,
title = "Total respiratory support with tidal flow extracorporeal circulation in adult sheep",
abstract = "A novel pressure gated tidal flow extracorporeal circulation (TF ECC) device was developed, and it was hypothesized that it could provide total respiratory support in apneic adult sheep without adverse hemodynamic or cardiac effects. The circuit consisted of a single lumen cannula, computer driven tubing occluders gated by circuit pressure, a nonocclusive peristaltic blood pump, a spiral coiled membrane lung, and a heat exchanger. Six paralyzed, anesthetized adult sheep were instrumented and TF ECC was instituted via cannulation of the right atrium. Total respiratory support was provided by the circuit during an apneic period of 6 hours. Echocardiography was performed with the animal instrumented (baseline) and after 2 hours of TF ECC. Circuit blood tidal volume was 172.6 ± 18.0 cc, resulting in a TF ECC flow of 71.1 ± 10.1 cc/kg/min. At the end of the study period, P(a)Co2 was 35.5 ± 7.6 mmHg, P(a)O2 was 91.2 ± 30.6 mmHg, and pulmonary artery oxygen saturation (S(PA)O2)) was 95 ± 5{\%}. Hemodynamic stability was maintained with no significant differences at baseline and after 6 hours in mean arterial pressure, mean pulmonary artery pressure, or heart rate noted. Echocardiographic evaluation showed preserved fractional shortening of the left ventricular (LV) septal-lateral dimension (baseline 32.4 ± 11.4 {\%}; 2 hours 34.8 ± 8.4 {\%}). This study demonstrates TF ECC provides total respiratory support without adverse hemodynamic effects, and preserved LV function.",
author = "Srinivas Kolla and Stefania Crotti and Lee, {W. Anthony} and Gargulinski, {Matthew J.} and Thomas Lewandowski and David Bach and Hirschl, {Ronald B.} and Bartlett, {Robert H.}",
year = "1997",
month = "9",
language = "English",
volume = "43",
journal = "ASAIO Journal",
issn = "0162-1432",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - Total respiratory support with tidal flow extracorporeal circulation in adult sheep

AU - Kolla, Srinivas

AU - Crotti, Stefania

AU - Lee, W. Anthony

AU - Gargulinski, Matthew J.

AU - Lewandowski, Thomas

AU - Bach, David

AU - Hirschl, Ronald B.

AU - Bartlett, Robert H.

PY - 1997/9

Y1 - 1997/9

N2 - A novel pressure gated tidal flow extracorporeal circulation (TF ECC) device was developed, and it was hypothesized that it could provide total respiratory support in apneic adult sheep without adverse hemodynamic or cardiac effects. The circuit consisted of a single lumen cannula, computer driven tubing occluders gated by circuit pressure, a nonocclusive peristaltic blood pump, a spiral coiled membrane lung, and a heat exchanger. Six paralyzed, anesthetized adult sheep were instrumented and TF ECC was instituted via cannulation of the right atrium. Total respiratory support was provided by the circuit during an apneic period of 6 hours. Echocardiography was performed with the animal instrumented (baseline) and after 2 hours of TF ECC. Circuit blood tidal volume was 172.6 ± 18.0 cc, resulting in a TF ECC flow of 71.1 ± 10.1 cc/kg/min. At the end of the study period, P(a)Co2 was 35.5 ± 7.6 mmHg, P(a)O2 was 91.2 ± 30.6 mmHg, and pulmonary artery oxygen saturation (S(PA)O2)) was 95 ± 5%. Hemodynamic stability was maintained with no significant differences at baseline and after 6 hours in mean arterial pressure, mean pulmonary artery pressure, or heart rate noted. Echocardiographic evaluation showed preserved fractional shortening of the left ventricular (LV) septal-lateral dimension (baseline 32.4 ± 11.4 %; 2 hours 34.8 ± 8.4 %). This study demonstrates TF ECC provides total respiratory support without adverse hemodynamic effects, and preserved LV function.

AB - A novel pressure gated tidal flow extracorporeal circulation (TF ECC) device was developed, and it was hypothesized that it could provide total respiratory support in apneic adult sheep without adverse hemodynamic or cardiac effects. The circuit consisted of a single lumen cannula, computer driven tubing occluders gated by circuit pressure, a nonocclusive peristaltic blood pump, a spiral coiled membrane lung, and a heat exchanger. Six paralyzed, anesthetized adult sheep were instrumented and TF ECC was instituted via cannulation of the right atrium. Total respiratory support was provided by the circuit during an apneic period of 6 hours. Echocardiography was performed with the animal instrumented (baseline) and after 2 hours of TF ECC. Circuit blood tidal volume was 172.6 ± 18.0 cc, resulting in a TF ECC flow of 71.1 ± 10.1 cc/kg/min. At the end of the study period, P(a)Co2 was 35.5 ± 7.6 mmHg, P(a)O2 was 91.2 ± 30.6 mmHg, and pulmonary artery oxygen saturation (S(PA)O2)) was 95 ± 5%. Hemodynamic stability was maintained with no significant differences at baseline and after 6 hours in mean arterial pressure, mean pulmonary artery pressure, or heart rate noted. Echocardiographic evaluation showed preserved fractional shortening of the left ventricular (LV) septal-lateral dimension (baseline 32.4 ± 11.4 %; 2 hours 34.8 ± 8.4 %). This study demonstrates TF ECC provides total respiratory support without adverse hemodynamic effects, and preserved LV function.