Use of INR calibrator plasmas in the routine coagulation laboratory: A study of two thrombolastin reagents

Annalisa Fattorini, Elisabetta Pattarini, Silvana Viganò, Luciano Crippa, Armando D'Angelo

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

INR values may be either calculated with the ISI values supplied by thromboplastin manufacturers or are directly extrapolated from certified INR calibrator plasmas. We tested the principle of local INR calibration using INR calibrator plasmas (PT-Multi Calibrator, Siemens), two thromboplastin reagents (Neoplastin Plus, rabbit brain, Stago, coagulometer-specific ISI 1.31, and Innovin, recombinant human tissue factor, Siemens) and the same coagulometer (STA-R, Stago) in 100 patients on warfarin. Using a ISI value of 0.77 with Tomenson correction for Innovin (correction factor = 1.09), INR values of patients were similar with the two reagents, with a bias of 0.03 INR units and no significant regression of the difference over the average INR by method comparison analysis. With the INR calibrator plasmas, INR values with Neoplastin Plus were lower than Innovin values with an average bias of 0.39 INR units and a significant regression of the difference over the average INR (r = - 0.91). Significant bias (0.16 INR units, p <0.00001) and regression (r = - 0.77) was also observed by comparison of Neoplastin Plus INRs with Innovin calibrated INRs. Based on a therapeutic INR interval of 2.0 to 3.5, discordance in warfarin dosing was approximately 3 times higher with INR calibration (27% vs 11%). Because of non commutability with fresh plasma samples, local INR calibration with lyophilized calibrator plasmas may not be valid for some reagent-instrument combinations.

Original languageEnglish
JournalThrombosis Research
Volume130
Issue number3
DOIs
Publication statusPublished - Sep 2012

Fingerprint

International Normalized Ratio
Thromboplastin
Calibration
Warfarin

Keywords

  • Commutability
  • INR calibrator plasmas
  • ISI
  • Local INR calibration
  • MNPT

ASJC Scopus subject areas

  • Hematology

Cite this

Use of INR calibrator plasmas in the routine coagulation laboratory : A study of two thrombolastin reagents. / Fattorini, Annalisa; Pattarini, Elisabetta; Viganò, Silvana; Crippa, Luciano; D'Angelo, Armando.

In: Thrombosis Research, Vol. 130, No. 3, 09.2012.

Research output: Contribution to journalArticle

Fattorini, Annalisa ; Pattarini, Elisabetta ; Viganò, Silvana ; Crippa, Luciano ; D'Angelo, Armando. / Use of INR calibrator plasmas in the routine coagulation laboratory : A study of two thrombolastin reagents. In: Thrombosis Research. 2012 ; Vol. 130, No. 3.
@article{9b6a74312ed74f8cbc154def2748a813,
title = "Use of INR calibrator plasmas in the routine coagulation laboratory: A study of two thrombolastin reagents",
abstract = "INR values may be either calculated with the ISI values supplied by thromboplastin manufacturers or are directly extrapolated from certified INR calibrator plasmas. We tested the principle of local INR calibration using INR calibrator plasmas (PT-Multi Calibrator, Siemens), two thromboplastin reagents (Neoplastin Plus, rabbit brain, Stago, coagulometer-specific ISI 1.31, and Innovin, recombinant human tissue factor, Siemens) and the same coagulometer (STA-R, Stago) in 100 patients on warfarin. Using a ISI value of 0.77 with Tomenson correction for Innovin (correction factor = 1.09), INR values of patients were similar with the two reagents, with a bias of 0.03 INR units and no significant regression of the difference over the average INR by method comparison analysis. With the INR calibrator plasmas, INR values with Neoplastin Plus were lower than Innovin values with an average bias of 0.39 INR units and a significant regression of the difference over the average INR (r = - 0.91). Significant bias (0.16 INR units, p <0.00001) and regression (r = - 0.77) was also observed by comparison of Neoplastin Plus INRs with Innovin calibrated INRs. Based on a therapeutic INR interval of 2.0 to 3.5, discordance in warfarin dosing was approximately 3 times higher with INR calibration (27{\%} vs 11{\%}). Because of non commutability with fresh plasma samples, local INR calibration with lyophilized calibrator plasmas may not be valid for some reagent-instrument combinations.",
keywords = "Commutability, INR calibrator plasmas, ISI, Local INR calibration, MNPT",
author = "Annalisa Fattorini and Elisabetta Pattarini and Silvana Vigan{\`o} and Luciano Crippa and Armando D'Angelo",
year = "2012",
month = "9",
doi = "10.1016/j.thromres.2012.05.015",
language = "English",
volume = "130",
journal = "Thrombosis Research",
issn = "0049-3848",
publisher = "Elsevier Limited",
number = "3",

}

TY - JOUR

T1 - Use of INR calibrator plasmas in the routine coagulation laboratory

T2 - A study of two thrombolastin reagents

AU - Fattorini, Annalisa

AU - Pattarini, Elisabetta

AU - Viganò, Silvana

AU - Crippa, Luciano

AU - D'Angelo, Armando

PY - 2012/9

Y1 - 2012/9

N2 - INR values may be either calculated with the ISI values supplied by thromboplastin manufacturers or are directly extrapolated from certified INR calibrator plasmas. We tested the principle of local INR calibration using INR calibrator plasmas (PT-Multi Calibrator, Siemens), two thromboplastin reagents (Neoplastin Plus, rabbit brain, Stago, coagulometer-specific ISI 1.31, and Innovin, recombinant human tissue factor, Siemens) and the same coagulometer (STA-R, Stago) in 100 patients on warfarin. Using a ISI value of 0.77 with Tomenson correction for Innovin (correction factor = 1.09), INR values of patients were similar with the two reagents, with a bias of 0.03 INR units and no significant regression of the difference over the average INR by method comparison analysis. With the INR calibrator plasmas, INR values with Neoplastin Plus were lower than Innovin values with an average bias of 0.39 INR units and a significant regression of the difference over the average INR (r = - 0.91). Significant bias (0.16 INR units, p <0.00001) and regression (r = - 0.77) was also observed by comparison of Neoplastin Plus INRs with Innovin calibrated INRs. Based on a therapeutic INR interval of 2.0 to 3.5, discordance in warfarin dosing was approximately 3 times higher with INR calibration (27% vs 11%). Because of non commutability with fresh plasma samples, local INR calibration with lyophilized calibrator plasmas may not be valid for some reagent-instrument combinations.

AB - INR values may be either calculated with the ISI values supplied by thromboplastin manufacturers or are directly extrapolated from certified INR calibrator plasmas. We tested the principle of local INR calibration using INR calibrator plasmas (PT-Multi Calibrator, Siemens), two thromboplastin reagents (Neoplastin Plus, rabbit brain, Stago, coagulometer-specific ISI 1.31, and Innovin, recombinant human tissue factor, Siemens) and the same coagulometer (STA-R, Stago) in 100 patients on warfarin. Using a ISI value of 0.77 with Tomenson correction for Innovin (correction factor = 1.09), INR values of patients were similar with the two reagents, with a bias of 0.03 INR units and no significant regression of the difference over the average INR by method comparison analysis. With the INR calibrator plasmas, INR values with Neoplastin Plus were lower than Innovin values with an average bias of 0.39 INR units and a significant regression of the difference over the average INR (r = - 0.91). Significant bias (0.16 INR units, p <0.00001) and regression (r = - 0.77) was also observed by comparison of Neoplastin Plus INRs with Innovin calibrated INRs. Based on a therapeutic INR interval of 2.0 to 3.5, discordance in warfarin dosing was approximately 3 times higher with INR calibration (27% vs 11%). Because of non commutability with fresh plasma samples, local INR calibration with lyophilized calibrator plasmas may not be valid for some reagent-instrument combinations.

KW - Commutability

KW - INR calibrator plasmas

KW - ISI

KW - Local INR calibration

KW - MNPT

UR - http://www.scopus.com/inward/record.url?scp=84865251529&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84865251529&partnerID=8YFLogxK

U2 - 10.1016/j.thromres.2012.05.015

DO - 10.1016/j.thromres.2012.05.015

M3 - Article

C2 - 22682443

AN - SCOPUS:84865251529

VL - 130

JO - Thrombosis Research

JF - Thrombosis Research

SN - 0049-3848

IS - 3

ER -