Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system

E. Mastella, S. Vigorito, E. Rondi, G. Piperno, A. Ferrari, Elena Strata, Daniela Rozza, B. A. Jereczek-Fossa, F. Cattani

Research output: Contribution to journalArticle

Abstract

Purpose: To evaluate the geometric and dosimetric accuracies of the CyberKnife Synchrony respiratory tracking system (RTS) and to validate a method for pretreatment patient-specific delivery quality assurance (DQA). Methods: An EasyCube phantom was mounted on the ExacTrac gating phantom, which can move along the superior-inferior (SI) axis of a patient to simulate a moving target. The authors compared dynamic and static measurements. For each case, a Gafchromic EBT3 film was positioned between two slabs of the EasyCube, while a PinPoint ionization chamber was placed in the appropriate space. There were three steps to their evaluation: (1) the field size, the penumbra, and the symmetry of six secondary collimators were measured along the two main orthogonal axes. Dynamic measurements with deliberately simulated errors were also taken. (2) The delivered dose distributions (from step 1) were compared with the planned ones, using the gamma analysis method. The local gamma passing rates were evaluated using three acceptance criteria: 3% local dose difference (LDD)/3 mm, 2%LDD/2 mm, and 3%LDD/1 mm. (3) The DQA plans for six clinical patients were irradiated in different dynamic conditions, to give a total of 19 cases. The measured and planned dose distributions were evaluated with the same gamma-index criteria used in step 2 and the measured chamber doses were compared with the planned mean doses in the sensitive volume of the chamber. Results: (1) A very slight enlargement of the field size and of the penumbra was observed in the SI direction (on average

Original languageEnglish
Pages (from-to)4565-4574
Number of pages10
JournalMedical Physics
Volume43
Issue number8
DOIs
Publication statusPublished - Aug 1 2016

Keywords

  • CyberKnife Synchrony system
  • delivery quality assurance
  • Gafchromic EBT3 film
  • multichannel dosimetry
  • stereotactic body radiation therapy

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint Dive into the research topics of 'Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system'. Together they form a unique fingerprint.

  • Cite this