Validation of a radial diffusive sampler for measuring occupational exposure to 1,3-butadiene

Mariella Carrieri, Giovanni B. Bartolucci, Enrico Paci, Paolo Sacco, Daniela Pigini, Laura Zaratin, Danilo Cottica, Maria L. Scapellato, Giovanna Tranfo

Research output: Contribution to journalArticlepeer-review

Abstract

1,3-Butadiene (BD) is a major industrial chemical used in the manufacture of rubbers and latexes; it is also a ubiquitous environmental pollutant whose major source is traffic. Occupational exposure to (BD) can occur both during its production and during its use as a raw material. The objective of the study was the laboratory and field validation of a new diffusive sampler for BD. The nominal sampling rate of the Radiello® diffusive sampler filled with Carbopack X is 30.5cm3/min, at 0.177mg/m3, 20°C and 50% relative humidity (RH), for an 8-h exposure time. A model can be used for calculating the sampling rate as a function of temperature, time and RH. The concentration does not affect the sampling rate above 30μg/m3. The measurement uncertainty (k=2), calculated both by laboratory data and by field comparison according to International Standard Organization (ISO) 13752, satisfies the EN 482:2006 requirement for measurements between 0.1 and 0.5 times the threshold limit value-time weighted average (TLV-TWA) (uncertainty®) in comparison with active samplers. The BD exposure levels detected for the exposed subjects were low (mean 0.059, range 3) but higher than the controls levels, all below 0.010mg/m3. The comparison between diffusive and active (pumped) air sampling showed a good correlation, with no systematic deviation from the ideal values of the intercept and slope of the optimized regression line. The concentrations of two biomarkers were also determined on urine samples, collected at the end of the work-shift: unchanged BD, by GC-MS, and the metabolite dihydroxybutylmercapturic acid (DHBMA), by HPLC-MS/MS. The urinary excretion of the biomarkers was on average higher in the exposed group (urinary BD: mean 8.8, range ®) appears to be a reliable method for the assessment of exposure to low levels of airborne BD and a convenient alternative to the conventional active sampling.

Original languageEnglish
Pages (from-to)114-120
Number of pages7
JournalJournal of Chromatography A
Volume1353
DOIs
Publication statusPublished - Aug 1 2014

Keywords

  • 1,3-Butadiene
  • Biological monitoring
  • Diffusive sampling
  • Environmental monitoring
  • Occupational exposure

ASJC Scopus subject areas

  • Analytical Chemistry
  • Organic Chemistry
  • Biochemistry
  • Medicine(all)

Fingerprint Dive into the research topics of 'Validation of a radial diffusive sampler for measuring occupational exposure to 1,3-butadiene'. Together they form a unique fingerprint.

Cite this