N2 - In the last three decades protocols for the validation of blood pressure measuring devices have been developed by the US Association for the Advancement of Medical Instrumentation, the British Hypertension Society, the German Hypertension League, the European Society of Hypertension Working Group on blood pressure Monitoring and the International Organization for Standardization. The European Society of Hypertension International Protocol required much smaller sample size than the other protocols, aiming to reduce the time, resources and cost of validation studies and thereby increase the number of validated devices. Given its specifications, the European Society of Hypertension International Protocol was adequate for 'high- and low-accuracy' devices, yet assessment of 'moderate accuracy' devices had high uncertainty with resultant high rate of device failure. Thus, devices validated using the European Society of Hypertension International Protocol should be considered to be as accurate as those validated with the previous Association for the Advancement of Medical Instrumentation or British Hypertension Society protocols. However, the European Society of Hypertension International Protocol did not allow subgroup evaluation (arm sizes, special populations, etc). The mission of the European Society of Hypertension International Protocol to promote the concept of validation has been well achieved, as almost double studies have been published using it than all the other protocols together. However, the maintenance of different validation protocols is confusing and therefore experts from the Association for the Advancement of Medical Instrumentation, European Society of Hypertension International Protocol and International Organization for Standardization have now developed the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) as the recommended 21st-century procedure for worldwide application. The European Society of Hypertension Working Group has published a practical guide for using the Universal Standard. It is in the interests of all scientific bodies to propagate the Universal Standard and ensure its wide implementation.
AB - In the last three decades protocols for the validation of blood pressure measuring devices have been developed by the US Association for the Advancement of Medical Instrumentation, the British Hypertension Society, the German Hypertension League, the European Society of Hypertension Working Group on blood pressure Monitoring and the International Organization for Standardization. The European Society of Hypertension International Protocol required much smaller sample size than the other protocols, aiming to reduce the time, resources and cost of validation studies and thereby increase the number of validated devices. Given its specifications, the European Society of Hypertension International Protocol was adequate for 'high- and low-accuracy' devices, yet assessment of 'moderate accuracy' devices had high uncertainty with resultant high rate of device failure. Thus, devices validated using the European Society of Hypertension International Protocol should be considered to be as accurate as those validated with the previous Association for the Advancement of Medical Instrumentation or British Hypertension Society protocols. However, the European Society of Hypertension International Protocol did not allow subgroup evaluation (arm sizes, special populations, etc). The mission of the European Society of Hypertension International Protocol to promote the concept of validation has been well achieved, as almost double studies have been published using it than all the other protocols together. However, the maintenance of different validation protocols is confusing and therefore experts from the Association for the Advancement of Medical Instrumentation, European Society of Hypertension International Protocol and International Organization for Standardization have now developed the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) as the recommended 21st-century procedure for worldwide application. The European Society of Hypertension Working Group has published a practical guide for using the Universal Standard. It is in the interests of all scientific bodies to propagate the Universal Standard and ensure its wide implementation.
UR - http://www.scopus.com/inward/record.url?scp=85069274149&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85069274149&partnerID=8YFLogxK
U2 - 10.1097/MBP.0000000000000391
DO - 10.1097/MBP.0000000000000391
M3 - Article
C2 - 31116156
VL - 24
SP - 163
EP - 166
JO - Blood Pressure Monitoring
JF - Blood Pressure Monitoring
SN - 1359-5237
IS - 4
ER -